CH ENGR C125 Study Guide - Quiz Guide: Theoretical Plate, Capacity Factor

88 views1 pages

Document Summary

Compound x: the following data applies to a column for liquid chromatography. Length of column is 35. 7 cm, the volume of the mobile and stationary phase is 1. 6 ml and 0. 134 ml, respectively. The column was run at a flow rate of 0. 5 ml/min. A mixture of 4 components (a, b, c, d) with the following chromatography data: 20. 6: component b and c are the closest component that may affect the separation. Determine the extent to which each molecule interacts with the stationary phase, the capacity factor (k") of b and c: determine the resolution (r) and the relative retention ( ) between components b and c. Was good separation achieved: text book problem no.