0
answers
0
watching
166
views
13 Dec 2019

Measuring the concentration of lead as (Pb2+) in paint samples using an atomic absorption instrument. A primary standard solution of aqueous Pb(NO3)2 is used to prepare standards for calibrating the instrument. Weighed out 0.2500 g of primary standard grade Pb(NO3)2 quantitatively transfers it to a container and dissolves it in DI water for a total solution volume of 250 mL. Various pieces of glassware are available including a 250 +/- 0.8mL graduated cylinder and 250 +/- 0.12 mL volumetric flask.

b. Estimate the uncertainty associated with preparing the solution using 1) the volumetric flask and 20 the graduated cylinder. Report the uncertainty as a range of concentrations (ppm Pb). Compare results.

c. What do these ranges of concentrations reveal about the difference in accuracy between the graduated cylinder and the volumetric flask? Explain using results from part b above, why volumetric flasks should always be chosen over graduated cylinders for analytical methods.

For unlimited access to Homework Help, a Homework+ subscription is required.

discord banner image
Join us on Discord
Chemistry Study Group
Join now

Related textbook solutions

Related questions

Weekly leaderboard

Start filling in the gaps now
Log in